MW-O101 Dissolved Oxygen Sensor

The MW-O101 sensor is a primary battery type dissolved oxygen water quality detection sensor. The oxygen molecules in the solution to be tested pass through the selective membrane of the sensor, and the corresponding reduction or oxidation reaction occurs on the cathode and anode inside the sensor, and a current signal is generated at the same time. The current is proportional to the dissolved oxygen concentration. oxygen concentration.

share

Related Products

No data

 
 

Sensor Features
Low power consumption, high precision, linear output, no need to replace the breathable membrane, no need to replace the electrolyte.

 

main application
It is widely used in dissolved oxygen water quality testing in laboratory scientific research, aquaculture, environmental protection and other fields.

 

Technical indicators

project

parameter

Measuring range

0~20 mg/L

temperature range

0~40 ℃

Minimum division value

0.01 mg/L

Response time (T90)

<20 s

Temperature compensation

without temperature compensation

Zero output (oxygen-free water)
(2% Na2SO3, 22℃)

< 0.06 μA

output

0.93 μA -1.33 μA in atmosphere at 22 °C

Measurement error

≤ ±0.1 mg/L

zero error

≤ 0.1 mg/L

Repeatability

≤ ±0.10 mg/L

stability

±0.03 mg/L

Sensitivity

0.6-1 mV/mg/L(O2)

output impedance

About 15 KΩ

interface

2 positive and negative connecting wires

Note: Product parameters are subject to change without prior notice

 

Instructions
Prepare 9 mg/L and 0 mg/L standard solutions for calibration. For the preparation method, please refer to the product use reminder card.
Connect the sensor interface end to the corresponding position on the module or instrument.
Remove the protective cover of the test end of the sensor and place it in the standard solution of 9 mg/L and 0 mg/L in turn for calibration.
After the calibration is completed, place the test end of the sensor in the liquid to be tested for testing.
Clean the sensor test end in time after use, and install the protective cover. Store at room temperature.